- CEM and SEM

- SEM 7M
Secondary electron mutiplier SEM-7M
As a part of scientific or industrial equipment, SEM-7M is intended to detect corpuscular and electromagnetic radiation. Secondary electron multiplier (SEM-7M) is used as a part of time-of-flight spectrometers. SEM-7M consists of two microchannel plates stacked with gap in chevron configuration. SEM body is metal-ceramic, soldered.
Secondary-electron multiplier SEM-7, intended for application in time-of-flight mass-spectrometry, is modernized. MCP 25-12F was replaced with MCP 25-10F, which allowed providing improved resolution and gain uniformity. Single electron amplitude resolution is 60%.
- Rise time 1,3÷1,4 ns.
- Time of flight 3÷3,2 ns.
| Physical Characteristics | ||
Parameter |
Dimension |
Value |
| Maximum diameter | mm | 50 |
| Height | mm | 10 |
| Input window area | cm2 | 5 |
| Key Characteristics | |||
Parameter |
Dimension |
Maximum |
Minimum |
| Gain | - | - | ×107 |
| Operating voltage | V | 2800 | - |
| Amplitude resolution | % | 70 | - |
| Dark count rate density | pulse/s-1× cm-2 | 3 | - |
| MCP current | µA | 25 | 1 |
| Weight | grammes | 25 | - |
Used types of MCP:
- 25-10 F with figured evaporated contact electrode;
- 25-12 F with figured evaporated contact electrode.
| MCP Parameters | ||
Parameter |
Dimension |
Value |
| Active diameter | mm | 25 |
| Channel diameter | um | 10; 12 |
| Bias angle | degree | 7; 13 |
MTTF not less than 1000 hrs.
There are 8 SEM modifications:
with MCP 25-10 F:
- with a grid in controlling electrode and with metal anode;
- with controlling electrode and with metal anode;
- with a grid in controlling electrode (without anode);
- with controlling electrode (without anode);
- with a grid in controlling electrode and with metal anode;
- with controlling electrode and with metal anode;
- with a grid in controlling electrode (without anode);
- with controlling electrode (without anode).